Ultrasound elastography in children — nice to have for scientific studies or arrived in clinical routine?
Molecular and Cellular Pediatrics volume 9, Article number: 11 (2022)
Abstract
Ultrasound elastography (USE) is a modality that in addition to fundamental B-mode, Doppler, and contrast-enhanced sonography is suitable to make qualitative and quantitative statements about the stiffness of tissues. Introduced more than 20 years ago in adults, USE becomes now a diagnostic tool also in children. The aim of this paper is to describe current available techniques for USE in children. The significance for routine use in children is shown, and further interesting applications are reported.
Introduction and history
Ultrasound is the central imaging method in pediatric diagnostics caused by the neglectable risk profile and high availability. In addition to fundamental B-mode sonography, Doppler modalities and contrast-enhanced ultrasound elastography (USE) were introduced more than 20 years ago into the clinical routine in adults increasing the diagnostic power of ultrasound examinations.
In medicine, palpation is a standard diagnostic tool in clinical examination and has always been used to differentiate benign and malignant lesions using the criteria of size, displaceability, compressibility, hardness, and consistency. Unfortunately, only superficial structures are accessible to palpation. Furthermore, palpation is a subjective method and strongly dependent on the experience of the physician. These limitations can be overcome with elastography, a procedure developed for both magnetic resonance (MR) and ultrasound. While MR elastography has so far not been able to establish in clinical routine due to considerable expenditure, USE plays an increasing role.
Elastography is used to measure noninvasively the internal tissue deformation that results after application of a force. According to the English ophthalmologist and physicist Thomas Young (1773–1829), elasticity is described using the modulus elasticity E, which is calculated from the ratio of tension to stretching of the tissue. The extent of relative changes in length when an external or internal force is applied depends on the elastic properties. Young’s modulus for diamond as the hardest material is 1220 GPa (Gigapascal), healthy soft tissue is between 0.5 and 70 kPa (Kilopascal), and healthy liver tissue is between 0.4 and 6 kPa [54].
Focal and diffuse findings of tissues could be described in more detail in terms of their elasticity. Starting with the evaluation of liver diseases in adults at the end of the 1990s, the horizon of elastography has subsequently expanded significantly. It was used to asses classic malignancies (e.g., breast cancer, prostate cancer) but also to evaluate superficial organs such as the thyroid gland and their focal lesions. Guidelines about the use of elastography in adults were published and updated by the European Federation of Societies for Ultrasound in Medicine and Biology (EFSUMB) and the World Federation for Ultrasound in Medicine and Biology (WFUMB) in adults for hepatic and non-hepatic applications [13, 22, 48]. Up to now, there are only few reports about features of elastography and techniques in children ([14], Mentzel 2020 [42]).
Technical background
The method of static elastography is based on the work of Ophir et al. who published elastograms based on the correlation analysis of tissue displacement under pressure [46]. At the beginning of the new century, dynamic elastography was developed superimposing elasticity values in color on the fundamental B-mode image in real time (real-time elastography, RTE). The strain image results from the different shifts of the echo frequency pattern under mechanical pressure (external compression, internal pulsation, breathing movement) on the analyzed tissue, with elastic tissues approaching the frequency peaks, while with stiff tissues, they remain at a constant distance. Rigid structures are usually shown in blue, intermediate deformable tissue parts are shown in yellow-green, and easily deformable tissue parts are shown in orange-red. Strain elastography provides a color-coded, qualitative result. The strain ratio can be used to estimate at least a ratio of the stiffness of the region of interest (ROI) to the surrounding area. Dynamic USE using shear waves is based on a focused impulse that propagates as a shock wave with a propagation speed of 1540 m/s into the tissue and leads to a compression (virtual touch). Compression is followed by the release of shear waves during the relaxation process. Shear waves propagate according to the rigidity of the environment. The speed of the shear waves (around 1–10 m/s) can be evaluated by a second ultrasonic pulse (acoustic radiation force impulse, ARFI) [21, 44]. Transient elastography (TE) with FibroScan© (Echosens, Paris, France) developed for liver elastography uses both ultrasound around 5 MHz and shear waves with 50 Hz generated by a defined external impulse vibrations. The result is given in kPa as an average of ten subsequent measurements [47]. Disadvantages of this well-standardized procedure are the lack of image control where the pulse is placed and the limitation in obese patients as well as in cases of ascites. ARFI is now widespread and available on most modern ultrasound devices with conventional transducers. A distinction has to be made between ARFI imaging as a qualitative method with representation of tissue deformation as a two-dimensional image and ARFI quantification (e.g., virtual touch quantification, VTQ, Siemens; Elast PQ, Philips) as quantitative tool (point shear wave elastography, pSWE). The measurement area is defined and placed as region of interest by means of B-mode imaging. In the past, ARFI has been expanded to include shear wave elastography (2D-SWE), in which the impulses are applied to a quantification box (Q-Box) at different tissue depths (e.g., supersonic shear imaging, SSI, Aix-en-Provence, France; virtual touch imaging quantification, VTIQ, Siemens, Germany; SWE, Philips, Netherlands; 2D-SWE, GE, USA; acoustic structure quantification, ASQ, Canon, Japan). The size of the evaluated region can be varied in real time, and the shear wave velocity (m/s) or the stiffness (kPa) can be specified [51] (Fig. 1).
The various techniques of USE can be compared among each other only with great caution. Measurements with different devices from various manufacturers, even data collected with different transducers on the same device, can hardly be related to one another. This must be taken into account for any follow-up checks at time intervals. Standardization of one’s own approach is required.
Clinical established applications
The liver is generally the organ most frequently evaluated by elastography. In addition to some reference value studies in children [41, 43], there are numerous clinical studies published on detecting and grading liver fibrosis as well as on diagnosis and treatment follow-up of biliary atresia in childhood ([16, 32]; Kim JR et al. [39];). Liver stiffness data can be expressed in kPa using TE or in m/s using pSWE or 2D-SWE. The software in modern ultrasound devices allows for conversion of m/s to Young’s modulus kPa for SWE. To assess diffuse liver changes, the right lobe of the liver is usually examined in segment 7. An intercostal access in a supine position with the right arm in extension gives more reproducible data than the evaluation of the left liver lobe [14]. Breath-hold minimizes measurement variability but can be difficult in children below the age of 5 years. Fast 2D-SWE evaluation with free-breathing gives results similar to breath-holding [37]. The liver should not move during the measurement. Region of interest is placed perpendicular to 2 cm subcapsular in the liver. It has proven useful to acquire 10 measurements and to use the median for further assessment [45]. In addition to age and gender, factors influencing liver stiffness are the positioning of the transducer and the region of interest as well as the blood flow to the liver, which depends on the time of last meal. The best reliability in the assessment of the liver stiffness is in examinations using 2D-SWE compared to pSWE and TE [38, 43]. There are promising USE studies for the evaluation of cystic fibrosis-related liver diseases (CFLD) [25]. A cutoff value of 1.27 m/s for CFLD was published achieving a specificity of 97% [9, 10]. For non-alcohol-related steatohepatitis (NASH), the combination of SWE and laboratory parameters has proven to be a biomarker of the severity of this entity [57]. In biliary atresia cases of post-Kasai liver fibrosis, ARFI can be used to assess liver fibrosis and to guide further treatment decisions like the urgency of liver transplantation [15, 29]. Further USE applications include the evaluation of liver pathology in alpha-1-antitrypsin deficiency, autoimmune hepatitis, veno-occlusive disease, hemochromatosis, Wilson’s disease, autosomal recessive polycystic kidney disease, and monitoring after liver transplantation [14, 28] (Fig. 2). USE plays a fixed role in the assessment of relevant liver pathologies and can be classified as a diagnostic component after clinical examination and standard liver laboratory prior to invasive liver biopsy [20, 24]. When evaluating liver elastography values, it should be borne in mind that there is a direct correlation with right-sided cardiac function. Every pathology of the right heart leads to pressure changes in the upstream vascular system, to the hepatic veins via the inferior vena cava. Due to the relatively coarse liver capsule, the changes in vascular pressure also lead to changes in rigidity. Each congestion ultimately leads to increased stiffness in the liver tissue. In children with congenital heart defects, liver stiffness can be used as a prognostic marker for cardiac events [23].
Other possible applications
While USE of the liver has meanwhile arrived in the clinical routine with children and adolescents, so far, all other applications are only poorly established.
Spleen
USE of the spleen is usually carried out in connection with examinations of portal hypertension in patients with liver fibrosis or cirrhosis and corresponding underlying diseases such as biliary atresia [63]. The examination is carried out in supine position in an intercostal incision from the lateral side. The region of interest is placed subcapsular at a depth of approximately 2 cm. Reference values given for pSWE (ARFI imaging) and TE can be used for the evaluation of spleen pathologies, e.g., evaluation of early signs of portal hypertension [9, 10, 30, 56].
Kidney
Caused by complex anatomy and given directionality in the parenchyma with cortex and medulla, the assessment of stiffness with USE is a challenge. Glomerulopathies in particular are a very interesting field for elastography [62]. Changes in the kidney transplant, acute and chronic inflammatory changes, should also have an influence of the parenchymal stiffness. Perfusion, hydration, and diuresis are factors which influence the elasticity of the kidney. Initial studies could show that the rigidity of the renal parenchyma increases with severe dilation [53]. Studies about the influence of vesicoureterorenal reflux on USE values are not without controversy [8]. So far, there are few studies on the use of USE in childhood. In the existing papers, the placement of the measurement region contradicts the standard customary in liver USE that the region of interest (ROI) has to be placed perpendicular to the ultrasound focus. Up to now, there are no standardized guidelines on the use of renal USE [33]. In our experience, the greatest reproducibility of measurement results can be achieved in prone position if, in the longitudinal view of the kidney, the measurement region is placed perpendicular to the focus in the middle of the parenchyma portion near to the transducer. Even then, both the cortex and medulla are visualized within the ROI.
Thyroid gland
In the thyroid gland, both the RTE as strain imaging and the SWE in its variations are used. Using RTE, a mechanical impulse administered by the examiner is given by the transducer. Alternatively, pulsation of the carotids can be used as stimulus. Scoring systems have been developed for adults which can be used for the estimation of nodules. Both strain imaging and shear wave elastography were used for autoimmune thyroiditis. In children, a significantly higher shear wave velocity and thus higher stiffness were found compared to children without pathology [4, 11]. Children with type I diabetes mellitus show a significant lower shear wave velocity [49]. USE can be used complementary to ultrasound in predicting thyroid nodules and may be helpful making a decision about fine needle or core biopsy [6, 31] (Fig. 3).
Lymph nodes
USE can be added in the ultrasound investigation of superficial lymph nodes using linear transducers. When assessing inflammatory changes in lymph nodes, strain elastography can be used to detect any meltdowns. The problem with USE is the lack of standardization of the examination. Various scoring systems are used to differentiate malignant from benign lymph nodes. Malignant transformed lymph nodes have a predominantly coarse pattern [64]. But, despite good results of USE, it cannot replace biopsy. So, it will be a challenge during the follow-up of children with former malignancy and new lymphadenopathy to differentiate between reactive lymph node enlargement, recurrence, or secondary malignancy [19] (Fig. 4).
Muscles
The mechanical properties of skeletal muscle include active and passive components related to function, growth, and metabolism. Activity can be characterized by electromyogram. The passive component can be analyzed using USE [61]. Strain imaging has proven itself in muscle elastography, even if there are only few studies on SWE. For strain elastography, a sufficient scoring system of 1–5 was introduced in children ranging from elastic-soft in predominantly red color (score 1), mostly soft in green with 25% red color (score 2), intermediate elastic in predominantly green (score 3) to predominantly hard in green with up to 25% blue (score 4), and finally stiff in blue (score 5) [5]. Using this scaling system, the muscles in children with spastic hemiparesis can be differentiated from healthy children, and, in perspective, the benefit of botulinum toxin therapy with corresponding loosening muscle effect can be monitored [59]. Using shear wave elastography on the gastrocnemic muscle, significantly higher passive stiffness was observed in children with unilateral cerebral palsy [7]. USE can become a valuable method in diagnosis and monitoring of musculoskeletal disorders in children [65].
Bowel
In the gastrointestinal tract, USE may be helpful for differentiating acute inflammation from chronic fibrous alterations of the bowel wall in children with chronic inflammatory bowel diseases (Fig. 5). So far, there are few pilot studies on scoring systems for children in order to estimate inflammatory activity [26]. According to a meta-analysis about 12 papers, USE may be a promising tool supporting evaluation of intestinal strictures in Crohn’s disease as well as the differentiation between fibrosis and inflammation in adults [52]. In addition to the sometimes limited compliance of infants, the frequently lively motility of the intestine which can lead to artifacts is problematic. There are initial studies on strain elastography and SWE in the assessment of acute appendicitis [2, 36].
Testes
The application and possible role of USE to testicular diseases are not well clarified. Up to now, there are few studies regarding the use of SWE for testicular torsion which showed that USE has potential to improve the differentiation between other testicular pathological conditions (e.g., tumor, abscess) and testicular torsion [55] (Fig. 6). In the central part of torsional testis, SWE showed no pathology caused by a mix of edema, hemorrhage, and necrosis. In the border of the twisted testis, SWE values were higher corresponding to increased tissue hardness. But, the twisting sign of the blood vessels (arteria and veins) is the golden standard for the clinical procedure and decision to go in the theater. Furthermore, elastography can be used to evaluate stiffness in undescended testis in comparison with the contralateral descended testes [17, 58] and in patients with microlithiasis who may show increased stiffness [3].
Brain
Initial reports showed that SWE of the periventricular parenchyma can be successfully performed in premature infants and children with hydrocephalus [12]. The extent to which elastography can be useful to determine increased intracranial pressure has to be verified through studies. Using 2D SWE in preterm infants, less stiffness was measured in the thalamus and in the periventricular parenchyma compared to full-term infants [1]. Anoxic brain injury can be quantified as an increase in stiffness caused by brain edema [12]. One possible application for USE in the future may be as a prognostic marker of neurodevelopmental outcome. Although first reference studies in terms of neonates exist, there are further reference studies necessary [34].
Lung
Lung ultrasound is imaging of artifacts. Intercostal B-mode imaging of the lung can differentiate A- and B-lines. The number of characteristic reverberation artifacts (B-lines) corresponds to the amount of extravascular lung water. Lung ultrasound surface shear wave elastography has proven itself for the assessment of pulmonary edema in patients with acute congestive heart failure and could show a reduction in shear wave speed suggesting a decrease in stiffness according to positive therapy effects [60]. Actually, USE was used in adults for the diagnosis and management of pneumonia in the COVID-19 pandemic [27]. Up to now, there are no reports about the application of lung USE in children. But, in the case of preterm babies, it could be possible to use elastography for monitoring neonatal respiratory distress syndrome. Further possible applications may be the evaluation of superficial focal lesions as well as monitoring pulmonary edema and interstitial lung disease with pleural/subpleural lesions [35].
Limitations and safety concern
Limitations lie in the inadequate standardization of the method. There is variability of measurement results using different devices and probes. In the liver, it could be shown that the results of USE are depending on a large number of variables. Compliance of the patient is necessary in most techniques of ultrasound elastography. The approval of ultrasound elastography is limited for use in abdominal, breast, thyroid, small parts, and musculoskeletal examinations. Shear wave elastography is approved by the Food and Drug Administration (FDA) in children and in adults.
The safety of ultrasound is monitored using thermal index (TI) associated with heating effects and mechanical index (MI) associated with cavitation effects. Both indices should be kept as low as possible. Potential risks arising from the energetic ultrasonic push pulse in elastography using acoustic displacement techniques (ARFI, SWE) must be considered [50]. The maximum temperature is at the focus of the evaluated region. The heating increases with the number of pulse sequences and scanning duration. Using a very strong ARFI, a temperature rise of 1.2 °C was estimated in a worst-case scenario (Fahey et al. 2015 [21]). But, in a 2D SWE study on mice, no histologically negative effects were found by SWE [40]. However, some temporary limited effects were observed if the scanning procedure lasted more than 30 min. As mentioned by the EFSUMB, the as low as reasonably achievable (ALARA) principle should be applied when setting the output for ARFI, and scanning time should be kept short [18]. Especially in case of vulnerable tissue (e.g., ovarian or testicular torsion, ischemic brain tissue), caution is necessary.
Conclusion
Ultrasound elastography which has been known for around 30 years was adopted into clinical routine in adulthood. The existing guidelines on adults may not be immediately transferrable to children [14]. Most of the experience with the use of various ultrasonic elastography methods comes from liver examinations — here, due to existing reference and cutoff value studies, elastography has now in many places become part of the routine in children and adolescents. But, there is still ample research potential for elastography in other organ systems.
Availability of data and materials
Not applicable.
Abbreviations
- ALARA:
-
As low as reasonably achievable
- ARFI:
-
Acoustic radiation force impulse
- B-mode:
-
Brightness mode
- cm:
-
Centimeter
- COVID-19:
-
Coronavirus SARS-CoV-2
- E:
-
Elasticity
- EFSUMB:
-
European Federation of Societies for Ultrasound in Medicine and Biology
- FDA:
-
Food and Drug Administration
- GPa:
-
Gigapascal
- kPa:
-
Kilopascal
- MI:
-
Mechanical index
- MR:
-
Magnetic resonance
- m/s:
-
Meter per second
- pSWE:
-
Point shear wave elastography
- ROI:
-
Region of interest
- RTE:
-
Real-time elastography
- SWE:
-
Shear wave elastography
- 2D SWE:
-
Two-dimensional shear wave elastography
- TE:
-
Transient elastography
- TI:
-
Thermal index
- USE:
-
Ultrasound elastography
- WFUMB:
-
World Federation of Ultrasound in Medicine and Biology
References
Albayrak E, Kasap T (2018) Evaluation of neonatal brain parenchyma using 2-dimensional shear wave elastography. J Ultrasound Med 37:959–967
Arslan H, Akdemir Z, Yavuz A (2018) Efficacy of strain elastography in diagnosis and staging of acute appendicitis in pediatric patients. Med Sci Monit 11:855–862
Aslan S, Bilgici MC, Saglam D et al (2018) The role of ARFI elastography to evaluate microstructural changes of patients with testicular microlithiasis. Acta Radiol 59:1517–1522. https://doi.org/10.1177/0284185118764213
Bakirtas Palabiyik B, Inci E, Papatya Cakir ED et al (2019) Evaluation of normal thyroid tissue and autoimmune thyroiditis in children using shear wave elastography. J Clin Paediatr Endocrinol 28:132–139
Berko NS, FitzGerald EF, Amaral TD et al (2014) Ultrasound elastography in children: establishing the normal range of muscle elasticity. Pediatr Radiol 44:158–163
Borysewicz-Sanczyk H, Dzieciol J, Sawicka B et al (2016) Practical application of elastography in the diagnosis of thyroid nodules in children and adolescents. Horm Res Paediatr 86:39–44. https://doi.org/10.1159/000447490
Boulard C, Gautheron V, Lapole T (2021) Mechanical properties of ankle joint and gastrocnemius muscle in spastic children with unilateral cerebral palsy measured with shear wave elastography. J Biomech. https://doi.org/10.1016/j.jbiomech.2021.110502
Bruno C, Caliari G, Zaffanello M et al (2013) Acoustic radiation force impulse (ARFI) in the evaluation of the renal parenchymal stiffness in paediatric patients with vesicoureteral reflux: preliminary results. Eur Radiol 23:3477–3484
Canas T, Macia A, Munoz-Codoceo RA et al (2015) Hepatic and splenic acoustic radiation force impulse shear wave velocity elastography in children with liver disease associated with cystic fibrosis. Biomed Res Int. https://doi.org/10.1155/2015/517369
Canas T, Fontanilla T, Miralles M et al (2015) Normal values of spleen stiffness in healthy children assessed by acoustic radiation force impulse imaging (ARFI): comparison between tow ultrasound systems. Pediatr Radiol 45:1316–1322
Cepeha CM, Paul C, Borlea A (2021) Shear-wave elastography – diagnostic value in children with chronic autoimmune thyroiditis. Diagnostics 11:248. https://doi.org/10.3390/diagnostics11020248
DeCampo D, Hwang M (2018) Characterizing the neonatal brain with ultrasound elastography. Ped Neurol 86:19–26
Dietrich CF, Bamber J, Berzigotti A et al (2017) EFSUMB guidelines and recommendations on the clinical use of liver ultrasound elastography, Update 2017. Ultraschall Med 38:e16–e47. https://doi.org/10.1055/s-0043-103952
Dietrich CF, Ferraioloi G, Sirli R et al (2019) General advice in ultrasound based elastography of pediatric patients. Med Ultrason 21:316–326. https://doi.org/10.11152/mu-2063
Dike PN, Mahmood N, Harpavat S (2021) Recent advances in the use of ultrasound and related techniques in diagnosing and predicting outcomes in biliary atresia. Curr Opin Pediatr 32. https://doi.org/10.1097/MOP.0000000000001048
Draijer LG, van Oosterhout JPM, Vali Y et al (2021) Diagnostic accuracy of fibrosis tests in children with non-alcoholic fatty liver disease: a systematic review. Liver Int 41:2087–2100. https://doi.org/10.1111/liv.14908
Durmaz MS, Sivri M, Sekmenli T et al (2018) Experience of using shear wave elastography imaging in evaluation of undescended testes in children: feasibility, reproducibility, and clinical potential. Ultrasound Q 34:206–212. https://doi.org/10.1097/RUQ.000000000000344
ECMUS tutorial: elastography – safety aspects. www.efsumb.org/ecmus. https://issuu.com/efsumb/docs/elastography_safety_aspects_2014
Elgendy A, Elhawary E, Shareef MM et al (2021) Ultrasound elastography in the diagnosis of malignant cervical lymphadenopathy in children: can it replace surgical biopsy? Eur J Pediatr Surg. https://doi.org/10.1055/s-0041-1729900
Engelmann G, Quader J, Teufel U et al (2017) Limitations and opportunities of non-invasive liver stiffness measurement in children. World J Hepatol 18:409–417. https://doi.org/10.4254/wjh.v9.i8.409
Fahey B, Nightingale KR, Nelson RC et al (2005) Acoustic radiation force impulse imaging of the abdomen: demonstration of feasibility and utility. Ultrasound in Med & Biol 31:1185–1198. https://doi.org/10.1016/j.ultrasmedbio.2005.05.004
Ferraioli G, Wong VE, Castera L et al (2018) Liver ultrasound elastography: an update to the World Federation for Ultrasound in Medicine and Biology guidelines and recommendations. Ultrasound Med Biol 44:2419–2440. https://doi.org/10.1016/j.ultrasmedbio.2018.07.008
Ferraioli G, Barr RG (2020) Ultrasound liver elastography beyond liver fibrosis assessment. World J Gastroenterol 26:3413–3420. https://doi.org/10.3748/wjg.v26i24.3413
Ferraioli G, Barr RG, Dillman JR (2021) Elastography for pediatric chronic liver disease. J Ultrasound Med 40:909–928. https://doi.org/10.1002/jum.15482
Fiege J, Mainz JG, Tabori H et al (2017) ARFI Elastografie der Leberbeteiligung bei zystischer Fibrose. Monatsschrift Kinderheilkd 165:888–894
Fufezan O, Asavoaie C, Tamas A et al (2015) Bowel elastography – a pilot study for developing an elastographic scoring system to evaluate disease activity in pediatric Crohn’s disease. Med Ultrason 17:422–430
Fusco P, Di Carlo S, Petroni GM et al (2021) Lung elastosonography for diagnosis and management of COVID-19 pneumonia. Minerva Anesthesiol 87:374–375. https://doi.org/10.23736/S0375-9393.20.15075
Galina P, Alexopoulou E, Mentessidou A (2021) Diagnostic accuracy of two-dimensional shear wave elastography in detecting hepatic fibrosis in children with autoimmune hepatitis, biliary atresia and other chronic liver diseases. Pediatr Radiol 51:1358–1368. https://doi.org/10.1007/s00247-020-04959-9
Hanquinet S, Courvoisier DS, Rougemont AL et al (2016) Acoustic radiation force impulse sonography in assessing children with biliary atresia for liver transplantation. Pediatr Radiol 46:1011–1016. https://doi.org/10.1007/s00247-016-3565-3
Hanquinet S, Habre C, Laurent M et al (2021) Acoustic radiation force impulse imaging: normal values of spleen stiffness in healthy children. Pediatr Radiol. https://doi.org/10.1007/s00247-021-05079-8
Hazem M, Zakaria OM, Daoud MYI et al (2021) Accuracy of shear wave elastography in characterization of thyroid nodules in children and adolescents. Insights Imaging 12:128. https://doi.org/10.1186/s13244-021-01074-7
Hwang JY, Yoon HM, Kim JR et al (2018) Diagnostic performance of transient elastography for liver fibrosis in children: a systematic review and meta-analysis. Am J Roentgenol 211:W257–W266. https://doi.org/10.2214/AJR.18.19535
Hwang J, Kim HW, Kim PH et al (2021) Technical performance of acoustic radiation force impulse imaging for measuring renal parenchymal stiffness: a systematic review and meta-analysis. J Ultrasound Med. https://doi.org/10.1002/jum.15654
Inigo EG, Salvador RL, Escrig R et al (2021) Quantitative evaluation of neonatal brain elasticity using shear wave elastography. J Ultrasound Med 40:795–804. https://doi.org/10.1002/jm.15464
Iovine E, Nenna R, Bloise S et al (2021) Lung ultrasound: its findings and new applications in neonatology and pediatric diseases. Diagnostics 11:652. https://doi.org/10.3390/diagnostics11040652
Isik IA, Ozkan MB (2021) Evaluation of the clinical effectiveness of shear wave elastography in pediatric cases with acute appendicitis. Ultrasound Q 37:173–177. https://doi.org/10.1097/RUQ.000000000000556
Jung C, Groth M, Petersen KU et al (2017) hepatic shear wave elastography in children under free-breathing and breath-hold conditions. Eur Radiol 27:5337–5343
Kim JR, Suh CH, Yoon HM et al (2018) The diagnostic performance of shear-wave elastography for liver fibrosis in children and adolescents: a systematic review and diagnostic meta-analysis. Eur Radiol 28:1175–1186. https://doi.org/10.1007/s00330-017-5078-3
Levitte S, Lee LW, Isaacson J et al (2021) Clinical use of shear-wave elastography for detecting liver fibrosis in children and adolescents with cystic fibrosis. Pediatr Radiol 51:1369–1377. https://doi.org/10.1007/s002247-021-05015
Li C, Zhang C, Li J et al (2016) Experimental study of the potential biological effects associated with 2-D shear wave elastography on the neonatal brain. Ultrasound Med Biol 42:1551–1559
Marginean CO, Melit LE, Ghiga DV et al (2020) Reference values of normal liver stiffness in healthy children by two methods: 2D shear wave and transient elastography. Sci Rep 29:7213. https://doi.org/10.1038/s41598-020-64320-w
Mentzel HJ (2020) Ultrasound elastography in childhood and adolescence. Monatschr Kinderheilkunde 168:427–433. https://doi.org/10.1007/s00112-020-00891-2
Mjelle AB, Mulabecirovic A, Havre RF et al (2019) Normal liver stiffness values in children: a comparison of three different elastography methods. J Pediatr Gastroenterol Nutr 68:706–712
Nightingale K, Mc Aleavey S, Trahey G (2003) Shear-wave generation using acoustic radiation force. In vivo and ex vivo results. Ultrasound Med Biol 29:1705–1713
Northern NA, Dillman JR, Trout AT (2019) Frequency of technical success of two-dimensional ultrasound shear wave elastography in a large pediatric and young adult cohort: a clinical effectiveness study. Pediatr Radiol 49:1025–1031
Ophir J, Cespedes I, Ponnekanti H et al (1991) Elastography: a quantitative method for imaging the elasticity of biological tissues. Ultrasound Imaging 13:111–134
Pawlus A, Sokolowska-Dabek D, Szymanska K et al (2015) Ultrsound elastography – review of techniques and its clinical applications in pediatrics – part 1. Adv Clin Exp Med 24:537–543. https://doi.org/10.17219/acem/34580
Saftoiu A, Gilja OD, Sidhu PS et al (2019) The EFSUMB guidelines and recommendations for the clinical practice on elastography in non-hepatic applications: update 2018. Ultraschall Med 40:425–453. https://doi.org/10.1055/a-0838-9937
Saglam D, Ceyhan Bilgici M, Kara C, e tal. (2017) Does type I diabetes melitus affect the shear wave velocity of the thyorid gland of children without autoimmune thyroiditis? Ultrasound Q 33:225–228
Schrenk F, Uhrik P, Uhrikova Z. Ultrasound elastography: review of techniques, clinical application, technical limitations, and safety considerations in neonatology. Acta Med Martiniana. 2020. https://doi.org/10.2478/acm-2020-0009.
Sigrist RMS, Liau J, Kaffas AE et al (2017) Ultrasound elastography: review of techniques and clinical applications. Theranostics 7:1303–1329
Slosarz D, Poniewierka E, Neubar K et al (2021) Ultrsound elastography in the assessment of the intestinal changes in inflammatory bowel disease – systematic review. J Clin Med 10:4044. https://doi.org/10.3390/jcm10184044
Sohn B, Kim MJ, Han SW et al (2014) Shear wave velocity measurements using acoustic radiation force impulse in young children with normal kidneys versus hydronephrotic kidneys. Ultrasonography 33:116–121
Stenzel M, Mentzel HJ (2014) Ultrasound elastography and contrast-enhanced sonography in infants, children, and adolescents. Eur J Radiol 82:1560–1569. https://doi.org/10.1016/j.ejrad.2014.06.007
Sun Z, Xie M, Xiang F et al (2015) Utility of real-time shear wave elastography in the assessment of testicular torsion. PLoS one:e0138523. https://doi.org/10.1371/journal.pone.0138523
Sutton H, Fitzpatrick E, Davenport M (2018) Transient elastography measurements of spleen stiffness as a predictor of clinically significant varices in children. J Pediatr Gastroenterol Nutr 67:446–451. https://doi.org/10.1097/MPG.0000000000002069
Thumar V, Squires JH, Spicer PJ et al (2018) Ultrasound elastography applications in pediatrics. Ultrasound Quarterly 34:199–205
Ucar AK, Alis D, Samanci C et al (2017) A prelminary study of shear wave elastography for the evaluation of unilateral palpable undescended testes. Eur J Radiol 86:248–251. https://doi.org/10.1016/j.ejrad.2016.11.026
Wentz E, Dieckmann A, Lehmann T et al (2019) Strain ultrasound elastography of muscles in healthy children and healthy adults. Fortsch Röntgenstr 191:1091–1098
Wiley BM, Zhou B, Pandompatam G et al (2021) Lung ultrasound surface wave elastography for assessing patients with pulmonary edema. IEEE Trans Biomed Eng 68:3417–3423. https://doi.org/10.1109/TBME.2021.3072891
Xiao Y, Wang C, Pan M et al (2021) Ultrasonic characterization of localized passive elastic properties of human pennate muscle with a single-probe setup. Ultrasonic. https://doi.org/10.1016/j.ultras2021.106512
Yogurtcuoglu B, Damar C (2021) Renal elastography measurements in children with acute glomerulonephritis. Ultrasonography. https://doi.org/10.14366/usg.20173
Yokoyama S, Ishigami M, Honda T et al (2019) Spleen stiffness by 2-D shear wave elastography is the most accurate predictor of high-risk esophagogastric varices in children with biliary atresia. Hepatol Research 49:1162–1168
Zakaria OM, Mousa A, AlSadhan R et al (2018) Reliabitlity of sonoelastography in predicting pediatric cervical lymph node malignancy. Pediatr Surg Int 34:885–890
Zaleska-Dorobisz U, Pawlus A, Szymanska K et al (2015) Ultrasound elastography – review of techniques and its clinical applications in pediatrics – part 2. Adv Clin Exp Med 24:725–730. https://doi.org/10.17219/acem/34581
Acknowledgements
Not applicable.
Funding
Open Access funding enabled and organized by Projekt DEAL.
Author information
Authors and Affiliations
Contributions
HJM, conceptualization, investigations, writing original draft, review, and editing. KG, investigations, validation, review, and editing. SG, investigations, analysis, review, and editing. PCK, investigations, analysis, review, and editing. MW, investigations, analysis, validation, review, and editing. The authors read and approved the final manuscript.
Authors’ information
Hans-Joachim Mentzel has been working for more than 20 years in pediatric radiology using all modalities of ultrasound imaging and approximately 10 years experience in ultrasound elastography. Katja Glutig has experience for 2 years in ultrasound elastography, Paul-Christian Krüger for 4 years, and Matthias Waginger for nearly 10 years.
Corresponding author
Ethics declarations
Ethics approval and consent to participate
Not applicable.
Consent for publication
Consent for publication was obtained.
Competing interests
The authors declare that they have no competing interests.
Additional information
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visithttp://creativecommons.org/licenses/by/4.0/.